H3531P compressed-air temperature and humidity sensor with alarms and Ethernet interface

H3531P compressed-air temperature and humidity sensor with alarms and Ethernet interface

code: H3531P
Compressed-air humidity, temperature transmitter with two relay outputs. T+RH probe from hard anodized duralumin for compressed air up to 25 bars with 1m cable. Cable lengths 2m or 4m available optionally.

Relative humidity, temperature probe on a cable. Measured values are also converted to other humidity interpretation: dew point temperature, absolute humidity, specific humidity, mixing ratio, specific enthalpy. Three two-state inputs.

 

Included in delivery: 

    • Traceable calibration certificate
    • Quick start manual
    • Free program TSensor for configuring of the transmitter is ready to Download
    • Free program SensorReader for logging values from one transmitter via Ethernet to a PC disk file is ready to Download
   
Features Technical Data Download Optional accessories We recommend

Features

Humidistat is designed for two-state control of e.g. heating, ventilation, humidifier, dehumidifier, etc.
Transmitter is equipped with two relay outputs for alarm indication or control of external devices. Each relay can be assigned to any measured or computed value. For each relay setting of delay, hysteresis, audible alarm is enabled.
Measured temperature and relative humidity is recalculated to other humidity interpretations - dew point temperature, absolute humidity, specific humidity, mixing ratio or specific enthalpy.
Parameters are easy adjustable from regulator keyboard or from the computer. Large dual line LCD for simultaneous display of temperature, relative humidity or other calculated humidity interpretation is an advantage.
State-of-the-art capacitive polymer sensor ensures excellent calibration long term stability, inertia against water and condensation. Transmitter is designed for use in non-aggressive environment.

Transmitter is also equipped with three binary inputs for detection of two-state events - e.g. water, smoke, glass break detection, door contact.

Easy creation of online data acquisition system from Ethernet sensors by using new low cost database software Comet

DBS Sensor Monitor

Online Ethernet Monitoring System

 

- see also Optional Accessories.

 

HUMIDISTAT APPLICATIONS

- temperature and humidity control
- monitoring of temperature, humidity:

  • building management and automation
  • server rooms
  • telecommunication devices
  • warehouses, glasshouses
  • manufacturers
  • museums, archives, galleries
  • air-conditioned rooms, HVAC
  • weather stations

 

MODES OF COMMUNICATION
Modbus TCP: Modbus TCP protocol enables to read measured values, set alarm limits, adjust the probe.
Telnet: Port 9999 enables to set alarm limits (lower, upper limits, hysteresis for measured values and time delay), e-mail addresses, SNMP addresses, probe description, refresh of www pages (10s to 65535s), set storing interval to history (10s to 65535s), enable each communication channel. Capacity of the history memory is 100 sets of temperature, humidity, pressure + computed values. Password protection of this port is enabled. Automatic IP address assignment from DHCP server is also enabled.
WWW pages: User selectable design of www pages enabling to display curve of measurement history. User can design the look of www pages and temperature, humidity, pressure + computed values history.
SNMP: It is possible to read actual values and alarm limits. In case of alarm creation warning message (trap) is sent to addresses defined by the user (maximum three addresses).
SOAP: Online transmitter enables to send actual measured data in the format of SOAP message to selected web server in preset interval 10-65535 s.

Online transmitter enables to monitor measured values. In case of exceeding of adjusted limits warning message can be sent to specified addresses. 

There are the following options of alarm indication:

ALARM INDICATION OPTIONS
E-mail: In case of alarm creation warning e-mail message is sent to addresses defined by the user (maximum three addresses) or via e-mail to SMS message. Basic SMTP autentization is supported.
SNMP: In case of exceeding of adjusted temperature, humidity, pressure + computed value limits alarm is activated and warning trap is sent to user specified IP addresses (maximum 3 addresses).
WWW pages: In case of exceeding of adjusted temperature, humidity, pressure + computed value limits active alarm is displayed on www page.
syslog: Online transmitter enables to send text messages to selected syslog server after different events appear. E.g. after transmitter restart, alarm activation, communication error with SNTP, after firmware change, after alarm termination, after communication error with SOAP server.

 

Synchronizing of time and record to www table and measured values history is enabled by SNTP via Internet.

 

For acquiring values from several transmitters, also third party software is possible to use. Tested functionality is with programs:

LabVIEW from National Instruments, by means of the free Modbus Library for LabVIEW.

TIRS.NET - transmitters are also directly supported by program TIRS.NET from Coral s.r.o. company. Please download demo enabling full work with transmitters.

Tested functionality is also with program ControlWeb

 

 Relative humidity range

Figure: Relative humidity operating range

 

Accuracy of dew point measurement

Figure: Accuracy of dew-point measurement

Technical Data

Technical parameters Value
Output Ethernet
Measured Value Relative Humidity + Temperature
Construction Type Compressed Air Humidity Up To 2.5MPa
Design Industrial
Temperature Measuring Range -30 to +105°C
Relay Output Yes
Two-State Input Yes
Lcd Display Yes
Relative humidity range 0 to 100%
Accuracy of relative humidity measurement ±2.5% relative humidity from 5 to 95% at 23°C
Accuracy of temperature output ±0.4°C
Resolution 0.1°C, 0.1%RH
Measuring interval 2s
Available temperature units degrees Celsius, degrees Fahrenheit
Computed values dew point, absolute humidity, specific humidity, mixing ratio, specific enthalpy
Accuracy and range of dew point temperature output - for more details see graphs in manual ±1.5°C at ambient temperature T<25°C and RH>30%
range -60 to +80°C
Accuracy and range of absolute humidity output ±1.5g/m3 at ambient temperature T < 25°C
range 0 to 400g/m3
Accuracy and range of specific humidity output ±2g/kg at ambient temperature T < 35°C
range 0 to 550g/kg
Accuracy and range of mixing ratio output ±2g/kg at ambient temperature T < 35°C
range 0 to 995g/kg
Accuracy and range of specific enthalpy output ±3kJ/kg at ambient temperature T < 25°C
range: 0 to 995 kJ/kg
Temperature compensation of the humidity sensor all temperature range
Temperature operating range -30 to +80°C
IP protection IP40
Number of relay outputs 2
Maximum switching voltage, current and power of relay output 50V, 2A, 60VA
Number of binary inputs 3
Signal for binary inputs dry contact, open collector or two-state voltage signal.
Inputs are not galvanically isolated.
Minimum pulse length at binary input 500ms
Low voltage level 0 to +0.5V
High voltage level +3.0 to +30V
Audible alarm built-in beeper - switchable
LAN connection RJ-45 connector, 10Base-T or 100Base-TX
Communication protocols WWW, ModbusTCP, SNMPv1, SOAP
Alarm protocols E-mail (SMTP authentication is supported), SNMP Trap, Syslog
Configuration T-Sensor, Telnet, keyboard
Temperature operating range of LCD display readable to operating temperature +70°C, it is recommended to switch OFF the LCD over +70°C
Filtering ability of sensor cover 0.025mm - filter with stainless steel mesh
Material of the probe duralumin with the black eloxal surface finish
Power 9 to 30Vdc, maximum consumption 1W
Power connector co-axial, diameter 5.5 x 2.1mm
Mechanical connection of probe G1/2 with O-Ring
Length of the probe cable 1m, 2m or 4m optionally
Dimensions 136 x 159 x 45mm (W x H x D), probe length 150mm, probe diameter 18mm
Weight approximately 460g - 1m probe
Warranty 2 years

  size
doc Instruction Manual 2.1 MB
doc Quick Start Manual 800 kB
doc Datasheet 1.5 MB
doc Declaration of Conformity 30 kB
doc MIB tables 250 kB

Program TSensor - free configuration program for transmitters and transducers, Code: TSensor more »
doc TSensor - free configuration program version 3.0.5.0 2.6 MB
doc TSensor - revision history 8 kB

Program SensorReader - free program for data read and store from one sensor, Code: SensorRead more »
doc SensorReader - free program for data recording from one sensor, version 2.0.1.0 1.3 MB

DBS Sensor Monitor - database program for data acquisition from Ethernet sensors P8xxx, Tx5xx, Hx5xx, Code: DBS more »
doc DBS Database Sensor Monitor version 2.3.0.1_II 55.2 MB

DBV Database Viewer for database programs DBL, DBM, DBS, Code: DBV more »
doc DBV Database Viewer 2.3.0.1 22.8 MB

Optional accessories

DBS Sensor Monitor - database program for data acquisition from Ethernet sensors P8xxx, Tx5xx, Hx5xx
code: DBS
Online monitoring system with sensors connected to Ethernet
DBV Database Viewer for database programs DBL, DBM, DBS
code: DBV
Low cost browser DBV Database Viewer enables several clients to view database from different places on network/internet.