Software

Thermo-hygrometer with external probe

Thermo-hygrometer with external probe

code: S3121
Data logger is designed for record of temperature and humidity. Values are stored to a non volatile electronic memory. Data transfer to the personal computer for further analysis is performed via USB, RS232, GSM or Ethernet interface by means of a proper communication adapter.

Included in delivery:

  • Data logger including the battery
  • Traceable calibration certificate with declared metrological traceability of etalons is based on requirements of EN ISO/IEC 17025 standard.
  • Quick start manual
  • Free basic Windows software is ready to download at folder Download

For communication with the PC must be from Optional accessories ordered USB adapter, LAN adapter or COM adapter or start/stop magnet if is needed to control logging the other way than directly from computer.

   
Features Technical Data Download Optional accessories We recommend

Features

Main applications of data logger:

  • Food and beverage industry (HACCP)
  • Pharmaceutical industry
  • Horticulture and cultivation of plants
  • HVAC (heating, ventilation, air conditioning, cooling)
  • Building and energy management
  • Research and development
  • Laboratories (GLP)
  • Warehouses
  • Technological processes
  • Museums, archives, galleries

Advantages of data logger:

  • included traceable calibration certificate from the manufacturer - Comet
  • fast data transfer to the PC (full memory for approximately 30s)
  • variability of connection to the computer - USB, RS232, Ethernet, GSM modem
  • permanent connection to the PC enabled, data is possible to download even during logging
  • large data memory 32000 values
  • large dual line display with special symbols, switchable
  • optional display of minimum and maximum measured values (reset of min/max memory from PC or by magnet)
  • dual level alarm is enabled for each channel, alarm is indicated by blinking of the value on the LCD display
  • two alarm modes are enabled: instant or with memory (detected alarm is indicated permanently till alarm memory is cleared from the PC)
  • robust watertight case
  • easy installation, locking enabled
  • low power consumption from battery, battery life up to 6 years
  • indication of remaining battery life, easy replacement of the battery
  • logging start/stop is enabled in several ways: at certain time and date programmed from computer, by delivered magnet
  • also special logging mode is enabled, when logging runs only, if some of measured values are out of adjusted alarm limits
  • each data logger is possible to describe with text of maximum 32 characters
  • each channel is possible to describe with text of maximum 16 characters
  • password protection is enabled to prevent unauthorized manipulation with logger

GSM communication

 

GSM COMMUNICATION - read more

  

relative humidity range

Figure: Relative humidity operating range

 

Accuracy of dew point measurement

Figure: Accuracy of dew-point measurement

Technical Data

Technical parameters Value
Measured Value Relative Humidity + Temperature
Construction Type With T+RH Probe On Cable
Temperature Measuring Range -30 to +105°C
Two-State Input No
Lcd Display Yes
Built-In Printer No
Operating temperature range -30 to +70°C
Measuring range of the probe temperature -30 to +105°C, humidity 0 to 100%RH
Accuracy of temperature measurement ±0.4°C from -30 to +80°C, ±0.5°C, over +80°C
Accuracy of humidity measurement ±2.5% RH from 5 to 95% at 23°C
Accuracy of dew point measurement - see details in graph 1.5 °C at ambient T < 25°C and RH>30% range -60 to +70 °C
Resolution of the reading 0.1°C, 0.1%RH
Real time clock year, leap year, month, day, hour, minute, second
Data logging interval adjustable from 10s to 24h (1minute to 24h in low-power mode)
Display and alarm refresh each 10 s (each 1 minute in low-power mode)
Total memory capacity 32000 temperature, humidity values in noncyclic logging mode
Data logging modes noncyclic - data logging stops after filling the memory
cyclic - after filling memory oldest data is overwritten by new
Dimensions without probe 93x64x29mm
Power Lithium battery 3,6V, size AA
Typical battery life in low-power mode (1 minute sampling) 6 years
Typical battery life in fast mode (10 second sampling) 2.5 years
Battery life in continuous on-line mode with interval 1 minute reduced to 70% of the above lives in fast mode
Battery life in continuous on-line mode with interval 10 second 1 year
IP protection IP67 - protected against influence of temporary immersion into water, RH+T sensors are covered with a stainless steel mesh with IP40 protection. Filtering ability 0,025mm
Weight including battery without probe 115g
Warranty 2 years

  size
doc Instruction Manual 560 kB
doc Datasheet 1.1 MB
doc Declaration of Conformity 60 kB

Program for Data Loggers Sxxxx, Rxxxx, Gxxxx, Dxxxx, Code: SWR004 more »
doc SWR004 Program for Data Loggers version 3.0.0.15 13.6 MB

DBL Logger Program - database software for Sxxxx, Rxxxx, Gxxxx data loggers, Code: DBL more »
doc DBL Logger Program version 3.0.0.17.II 61.8 MB

DBV Database Viewer for database programs DBL, DBM, DBS, Code: DBV more »
doc DBV Database Viewer 2.3.0.1 22.8 MB

Optional accessories

DBL Logger Program - database software for Sxxxx, Rxxxx, Gxxxx data loggers
code: DBL
Database program package.
DBV Database Viewer for database programs DBL, DBM, DBS
code: DBV
Low cost browser DBV Database Viewer enables several clients to view database from different places on network/internet.
Program for Data Loggers Sxxxx, Rxxxx, Gxxxx, Dxxxx
code: SWR004
Program for Windows for data loggers of Rxxxx, Sxxxx, Gxxxx, Dxxxx, ZV families.
LAN adapter with 0.5 meter cable
code: LP005
LAN adapter - external converter for Sxxxx, Rxxxx loggers communication with the PC via Ethernet, including ac/dc adapter 100 to 240Vac/5Vdc. Length of the cable 0.5 m. Protection IP65. Exceeding of adjusted limits is alarmed by sending e-mail message or SNMP trap.
S3121P Thermo-hygrometer for compressed air measurement
code: S3121P
Data logger is designed for record of temperature and humidity. Values are stored to a non volatile electronic memory. Data transfer to the personal computer for further analysis is performed via USB, RS232, GSM or Ethernet interface by means of a proper communication adapter.
Thermometer hygrometer data logger
code: S3120
Data logger is designed for record of temperature and humidity. Values are stored to a non volatile electronic memory. Data transfer to the personal computer for further analysis is performed via USB, RS232, GSM or Ethernet interface by means of a proper communication adapter.
Thermo-hygrometer + 2x 0-5Vdc voltage input with disp.
code: S3541
Data logger is designed for record of temperature and humidity. Values are stored to a non volatile electronic memory. Data transfer to the personal computer for further analysis is performed via USB, RS232, GSM or Ethernet interface by means of a proper communication adapter.
Thermo-hygrometer + external probe
code: R3121
Thermometer hygrometer is designed for record of temperature. Values are stored to a non volatile electronic memory. Data transfer to the personal computer for further analysis is performed via USB, RS232, GSM or Ethernet interface by means of a proper communication adapter.
D3121 Thermo-hygrometer
code: D3121
Thermometer, hygrometer is designed for measurement and datalogging of temperature, relative humidity into the internal non-volatile memory. Free PC program enables to transfer recorded data via RS232 serial link to the computer for evaluation. Data logger is connected to the PC through included communication cable only for data trasfer from the memory. During connection with the PC no measurement and record is enabled.